AE Techron

Showing 9–10 of 10 results

  • AE Techron Model T2000 Low-Frequency Conducted Susceptibility Transformer

    The T2000 Audio-Bandwidth Conducted Susceptibility Transformer was designed to meet or exceed the LF conducted susceptibility test requirements of DO 160 Section 18. The T2000 is used to apply the required test signals to the lines under test for ripple voltage tests and continuous or transient conducted immunity tests. It may also be used as a 2-to-1 step-down transformer. The secondary can support up to 40AP.

    The T2000 transformer supports up to 200W and is able to pass up to 40AP on the secondary. The turns ratio provides a 2-to-1 step down.

    The T2000 provides convenient input connectors via binding posts. Standard 0.75-inch spacing of binding posts allows use of standard plugs. Output is via Multi Contact 125A plugs.

    A thermal breaker protects the unit against over-current conditions on the primary. A 35 ARMS breaker protects the unit against over-current conditions on the secondary. A rugged, impact-resistant case and robust design protects the transformer from accidental damage.

    • Frequency response: 10 Hz to 250 kHz
    • Up to 200W audio power
    • Up to 40 Ap AC or DC secondary current
    • Two-to-one step-down
    • Output via Multi Contact 125A plugs
    • Testing  to DO 160 Section 16 and 18 standard
  • AE Techron Model T3700 Electric Field Susceptibility Transformer

    The T 3700 Electric-Field Susceptibility Transformer was designed to meet or exceed the voltage requirements of DO 160 Section 19 and Boeing D6-16050 Section 7.2 testing. It may also be used as a 1-to-37 step-up transformer for voltages up to 3700 VRMS.

    The T 3700 transformer is able to generate up to a 3700 VRMS potential. It is withstand tested to 135% of rated output to ensure a long, trouble-free life. The turns ratio provides a 1-to-37 step up to the secondary winding.

    The AE Techron  T 3700 provides convenient input connectors via touch-proof, four-way binding posts. Standard 0.75-inch spacing of binding posts allows use of standard plugs.

    A 1-amp fuse protects the unit against over-current conditions. A rugged, impact-resistant case and robust design protects the transformer from accidental damage.

    The T 3700’s voltage output meets or exceeds the requirements of DO 160 Section 19 and Boeing D6-16050-5 Section 7.2. It is for use by experienced staff.

    • Up to 3700VRMS
    • Robust withstand testing to 5,000 VDC
    • 1-to-37 step-up
    • Frequency range: 250 Hz to 35 kHz
    • Output via LGH high-voltage connector
    • Testing  to DO 160 Section 19 and Boeing D6-16050-5 Section 7.2 standards